Low radiation dose non-contrast cardiac CT: is it of value in the evaluation of mechanical aortic valve.
Abstract
BACKGROUND
Prosthetic bileaflet mechanical valve function has been traditionally evaluated using echocardiography and fluoroscopy. Multidetector
computed tomography (MDCT) is a novel technique for cardiac evaluation.
PURPOSE
To evaluate bileaflet mechanical aortic valves using a low-milliampere (mA), non-contrast MDCT protocol with a limited scan
range.
MATERIAL AND METHODS
Forty patients with a bileaflet mechanical aortic valve were evaluated using a non-contrast, low-mA, ECG-gated 64 MDCT protocol
with a limited scan range. MDCT findings of opening and closing valve angles were correlated to fluoroscopy and echocardiography.
Also, the valve visibility was evaluated on MDCT and fluoroscopy according to a 3-point grading scale.
RESULTS
The visualization score with the MDCT was significantly superior to the fluoroscopy (3 vs. 2.7). A strong correlation was
noted between the opening (r = 0.82) and closing (r = 0.96) valve angles with MDCT and fluoroscopy without a statistically
significant difference (P = 0.31 and 0.16, respectively). The mean effective radiation dose of the suggested protocol was
4 ± 0.5 mSv. Five valves were evaluated using transesophageal echocardiography because the valves were difficult to evaluate
with transthoracic echocardiography, and all of these valves were evaluated optimally with MDCT. A high-pressure gradient
was noted in nine valves, and the MDCT showed that seven of these valves inadequately opened, and two valves opened well,
which resulted in patient valve mismatch. Incomplete valve closure was noted in five valves, and the echocardiography showed
significant transvalvular regurgitation in all five valves.
CONCLUSION
MDCT can provide a precise measurement of valve function and can potentially evaluate high-pressure gradients and transvalvular
regurgitation.
Links
Authors
Bazeed MF, Moselhy MS, Rezk AI, Al-Murayeh MA
Institution
Department of Diagnostic Radiology, Faculty of Medicine, Mansoura University, Egypt. m_bazeed@yahoo.com
Source
Acta radiologica (Stockholm, Sweden : 1987) 53:4 2012 May 1 pg 389-93MeSH
AdultAortic Valve
Cardiac-Gated Imaging Techniques
Echocardiography, Transesophageal
Female
Fluoroscopy
Heart Valve Prosthesis
Humans
Male
Prospective Studies
Radiation Dosage
Radiographic Image Interpretation, Computer-Assisted
Tomography, X-Ray Computed
Pub Type(s)
Comparative StudyJournal Article
Language
eng
PubMed ID
22422269
Log In