Time-delayed summation as a means of improving resolution on fast rotating computed tomography systems.
Abstract
PURPOSE
Modern computed tomography (CT) systems are supporting increasingly fast rotation speeds, which are a prerequisite for fast
dynamic acquisition, e.g. in perfusion imaging, and for new modalities such as dedicated breast CT, where breathhold scanning
is indicated. However, not all detector technologies are supporting the high frame rates that are necessary to retain high
resolution for objects far away from the isocenter. Even on systems that would support a sufficiently high frame rate, the
necessary bandwidth of the data transfer from the rotating gantry stills remains challenging. The authors evaluated a pixel
shifting technique termed time-delayed summation (TDS) as a method of increasing resolution on fast rotating CT systems without
the need to increase the frame rate.
METHODS
In TDS mode, detector pixel values are shifted along rows during image acquisition to compensate for detector motion. In order
to fully exploit TDS, focal spot position control (FSC) was used in combination with TDS. FSC applies a counter movement to
the x-ray focal spot during image acquisition such that it is kept fixed in space. As a proof of concept, measurements were
performed on a prototype photon counting detector capable of TDS. The detector was mounted on a movable table and a gold wire
phantom was imaged with different TDS settings and detector velocities. Additionally, simulations of a broad range of TDS
and FSC settings on two different modalities, a clinical CT scanner and a breast CT scanner, and two different detector geometries,
flat and cylindrical, were performed to assess the gain in resolution and contrast in cylindrical water phantoms containing
a small wire at distances from the phantom center varied from 5% to 90% of the phantom radius. As figures of merit, the modulation
transfer function (MTF) at 10% and the maximum contrast were used and compared against the respective values when using step-and-shoot
acquisition, which means stopping the rotation when a projection image is acquired.
RESULTS
Measurements showed that detector movement and the resulting blurring of the wire projections were compensated to the expected
degree when using the appropriate number of TDS shifts per frame (TDS factor). Using simulations it was found that when using
the optimal TDS factor, over 90% of the resolution achieved in step-and-shot mode was reached for all investigated wire positions.
TDS showed better performance on a cylindrical detector that on the same system with a flat detector. TDS factors that were
deviating from the optimum by more than 1 shift led to a performance below that of standard continuous acquisition.
CONCLUSIONS
The findings of this study encourage the combined usage of TDS and FSC in systems that require fast rotation. The integration
of TDS in state-of-the-art x-ray detectors is feasible.
Links
Authors
Nowak T, Hupfer M, Althoff F, Brauweiler R, Eisa F, Steiding C, Kalender WA
Institution
Institute of Medical Physics, University of Erlangen-Nürnberg, Henkestraße 91, Erlangen, Germany. Tristan.Nowak@imp.uni-erlangen.de
Source
Medical physics 39:4 2012 Apr pg 2249-60MeSH
AlgorithmsHumans
Imaging, Three-Dimensional
Phantoms, Imaging
Radiographic Image Enhancement
Radiographic Image Interpretation, Computer-Assisted
Reproducibility of Results
Rotation
Sensitivity and Specificity
Time Factors
Tomography, X-Ray Computed
Pub Type(s)
Journal ArticleResearch Support, Non-U.S. Gov't
Language
eng
PubMed ID
22482646
Log In

