| Title | Comparison of the St. Petersburg phantom with a BOMAB phantom in the ORTEC StandFast whole body counter: a Monte Carlo simulation. | | Author(s) | Kramer GH, Capello K, Sung J | | Institution | Human Monitoring Laboratory, Radiation Surveillance and Health Assessment Division, Radiation Protection Bureau, 775 Brookfield Road, PL6302D1, Ottawa, Ontario, Canada K1A 1C1. gary_h_kramer@hc-sc.gc.ca | | Source | Health Phys 2008 May; 94(5 Suppl 2):S78-82. | | Abstract | Three sizes of the St. Petersburg phantom have been compared to six sizes of BOMAB phantoms measured by a virtual StandFast whole body counter using Monte Carlo simulations to investigate if the counting efficiencies are equivalent. This work shows that previously published data comparing the Reference Man sized phantom at 662 keV is supported; however, the simulations also show that the smaller sized St. Petersburg phantoms do not agree well with BOMAB phantoms. It is concluded that, compared with BOMAB phantoms, the St. Petersburg phantoms are system dependent and that they should be validated over a wide photon energy range against corresponding BOMAB phantoms prior to their use for calibrating whole body counters. | | Language | eng | | Pub Type(s) | Journal Article
| | PubMed ID | 18403961 |
|