Light scatter on the surface of AcrySof intraocular lenses: part I. Analysis of lenses retrieved from pseudophakic postmortem human eyes. Ophthalmic surgery, lasers & imaging : the official journal of the International Society for Imaging in the Eye [Ophthalmic Surg Lasers Imaging] Journal article

 
Yaguchi S, Nishihara H, Kambhiranond W, Stanley D, Apple DJ 
Light scatter on the surface of AcrySof intraocular lenses: part I. Analysis of lenses retrieved from pseudophakic postmortem human eyes. [Journal Article, Research Support, Non-U.S. Gov't]
Ophthalmic Surg Lasers Imaging 2008 May-Jun; 39(3):209-13.


BACKGROUND AND
OBJECTIVE: To investigate the cause of light scatter measured on the surface of AcrySof intraocular lenses (Alcon Laboratories, Inc., Fort Worth, TX) retrieved from pseudophakic postmortem human eyes.
MATERIALS AND METHODS: Ten intraocular lenses (Alcon AcrySofModel MA60BM) were retrieved postmortem and analyzed for light scatter before and after removal of surface-bound biofilms.
RESULTS: Six of the 10 lenses exhibited light scatter that was clearly above baseline levels. In these 6 lenses, both peak and average pixel density were reduced by approximately 80% after surface cleaning.
CONCLUSION: The current study demonstrates that a coating deposited in vivo on the lens surface is responsible for the light scatter observed when incident light is applied.