| Title | Cyclodextrin-MEEKC for the analysis of oxybutynin and its impurities. | | Author(s) | Giannini I, Orlandini S, Gotti R, Pinzauti S, Furlanetto S | | Institution | Department of Pharmaceutical Sciences, University of Florence, Via U. Schiff 6, 50019 Sesto Fiorentino, Florence, Italy. | | Source | Talanta 2009 Dec 15; 80(2):781-8. | | Abstract | The development of a cyclodextrin-MEEKC method for the analysis of oxybutynin and five related impurities is described. Experimental design strategies were applied in order to reach baseline separation of the compounds in a short analysis time. Mixture design made it possible to find the best composition for the microemulsion acting as pseudostationary phase, which was constituted by 89.1% 10mM borate buffer pH 9.2, 1.7% n-heptane, 9.2% SDS/n-butanol in 1:2 ratio. The addition of (2-hydroxypropyl)-beta-cyclodextrin to the background electrolyte was found to improve analysis performance. A Doehlert design, for the factors cyclodextrin concentration and voltage, was carried out and Derringer desirability function led to the identification of 18mM and 29kV as the optimal values. Applying the optimum conditions, separation of all the compounds, including the enantiomers of impurity 1, was obtained in less than 12min. The method was validated according to ICH guidelines for drug assay and determination of impurities and was applied to oxybutynin tablet analysis. | | Language | eng | | Pub Type(s) | Journal Article