[Dampness in an electric plug as a cause of electricity failure in an operation theatre].Ugeskr Laeger. 2000 Feb 07; 162(6):799-800.UL
Abstract
Two cases of electricity failure in an operation theatre during open heart surgery are discussed. The fuse for the patient monitor, ventilator, surgery instruments and heart lung machine was blown. Short-circuit was established because of humidity in the plug of the heater for fluid and blood. We recommend sealed or founded plugs and that anaesthesia equipment should not be used as an electrical supply for other electronic apparatus.
MeSH
Pub Type(s)
Case Reports
English Abstract
Journal Article
Language
dan
PubMed ID
10689957
Citation
Andersen, C, et al. "[Dampness in an Electric Plug as a Cause of Electricity Failure in an Operation Theatre]." Ugeskrift for Laeger, vol. 162, no. 6, 2000, pp. 799-800.
Andersen C, Pold R, Nielsen HD. [Dampness in an electric plug as a cause of electricity failure in an operation theatre]. Ugeskr Laeger. 2000;162(6):799-800.
Andersen, C., Pold, R., & Nielsen, H. D. (2000). [Dampness in an electric plug as a cause of electricity failure in an operation theatre]. Ugeskrift for Laeger, 162(6), 799-800.
Andersen C, Pold R, Nielsen HD. [Dampness in an Electric Plug as a Cause of Electricity Failure in an Operation Theatre]. Ugeskr Laeger. 2000 Feb 7;162(6):799-800. PubMed PMID: 10689957.
* Article titles in AMA citation format should be in sentence-case
TY - JOUR
T1 - [Dampness in an electric plug as a cause of electricity failure in an operation theatre].
AU - Andersen,C,
AU - Pold,R,
AU - Nielsen,H D,
PY - 2000/2/26/pubmed
PY - 2000/3/4/medline
PY - 2000/2/26/entrez
SP - 799
EP - 800
JF - Ugeskrift for laeger
JO - Ugeskr Laeger
VL - 162
IS - 6
N2 - Two cases of electricity failure in an operation theatre during open heart surgery are discussed. The fuse for the patient monitor, ventilator, surgery instruments and heart lung machine was blown. Short-circuit was established because of humidity in the plug of the heater for fluid and blood. We recommend sealed or founded plugs and that anaesthesia equipment should not be used as an electrical supply for other electronic apparatus.
SN - 0041-5782
UR - https://www.unboundmedicine.com/medline/citation/10689957/[Dampness_in_an_electric_plug_as_a_cause_of_electricity_failure_in_an_operation_theatre]_
DB - PRIME
DP - Unbound Medicine
ER -