Development of a flexible inflow cannula with titanium inflow tip for the NEDO biventricular assist device.
ASAIO J. 2004 Jul-Aug; 50(4):381-6.AJ

Abstract

A newly designed flexible inflow cannula has been developed for a biventricular assist device (BVAD). The inflow tip was designed for long-term ventricle drainage. Considering the anatomic differences between the right and left ventricles, a flexible inflow cannula, as a mass production model with a titanium tip, was developed and investigated in chronic BVAD animal experiments using the NEDO permanently implantable centrifugal blood pump. These flexible inflow cannulae consist of flexible plastic tubing with wire supported, titanium left and right tips. The right titanium inflow tip was designed with inner and outer stoppers to be inserted into the right ventricle wall and a beak shape to avoid sucking. The left tip was designed to be inserted into the left ventricle cavity. Five chronic BVAD bovine studies were performed to investigate the new beak shaped titanium tip. In these studies, the new beak shaped titanium tip and cannula prevented stenosis by a proliferation of pseudoneointimal, and supplied adequate blood flow to the pump without sucking. This newly designed inflow cannula with the beak shaped titanium tip was successfully implanted for 90 days.

Authors+Show Affiliations

Baylor College of Medicine, Houston, TX 77030, USA.No affiliation info availableNo affiliation info availableNo affiliation info availableNo affiliation info availableNo affiliation info availableNo affiliation info availableNo affiliation info availableNo affiliation info availableNo affiliation info available

Pub Type(s)

Evaluation Study
Journal Article
Research Support, Non-U.S. Gov't

Language

eng

PubMed ID

15307553

Citation

Watanabe, Kuniyoshi, et al. "Development of a Flexible Inflow Cannula With Titanium Inflow Tip for the NEDO Biventricular Assist Device." ASAIO Journal (American Society for Artificial Internal Organs : 1992), vol. 50, no. 4, 2004, pp. 381-6.
Watanabe K, Asai T, Ichikawa S, et al. Development of a flexible inflow cannula with titanium inflow tip for the NEDO biventricular assist device. ASAIO J. 2004;50(4):381-6.
Watanabe, K., Asai, T., Ichikawa, S., Motomura, T., Hata, A., Ito, S., Shinohara, T., Tsujimura, S., Oestmann, D. J., & Nosé, Y. (2004). Development of a flexible inflow cannula with titanium inflow tip for the NEDO biventricular assist device. ASAIO Journal (American Society for Artificial Internal Organs : 1992), 50(4), 381-6.
Watanabe K, et al. Development of a Flexible Inflow Cannula With Titanium Inflow Tip for the NEDO Biventricular Assist Device. ASAIO J. 2004 Jul-Aug;50(4):381-6. PubMed PMID: 15307553.
* Article titles in AMA citation format should be in sentence-case
TY - JOUR T1 - Development of a flexible inflow cannula with titanium inflow tip for the NEDO biventricular assist device. AU - Watanabe,Kuniyoshi, AU - Asai,Toshimasa, AU - Ichikawa,Seiji, AU - Motomura,Tadashi, AU - Hata,Atsushi, AU - Ito,Seiichi, AU - Shinohara,Toshiyuki, AU - Tsujimura,Shinichi, AU - Oestmann,Daniel J, AU - Nosé,Yukihiko, PY - 2004/8/17/pubmed PY - 2004/10/20/medline PY - 2004/8/17/entrez SP - 381 EP - 6 JF - ASAIO journal (American Society for Artificial Internal Organs : 1992) JO - ASAIO J VL - 50 IS - 4 N2 - A newly designed flexible inflow cannula has been developed for a biventricular assist device (BVAD). The inflow tip was designed for long-term ventricle drainage. Considering the anatomic differences between the right and left ventricles, a flexible inflow cannula, as a mass production model with a titanium tip, was developed and investigated in chronic BVAD animal experiments using the NEDO permanently implantable centrifugal blood pump. These flexible inflow cannulae consist of flexible plastic tubing with wire supported, titanium left and right tips. The right titanium inflow tip was designed with inner and outer stoppers to be inserted into the right ventricle wall and a beak shape to avoid sucking. The left tip was designed to be inserted into the left ventricle cavity. Five chronic BVAD bovine studies were performed to investigate the new beak shaped titanium tip. In these studies, the new beak shaped titanium tip and cannula prevented stenosis by a proliferation of pseudoneointimal, and supplied adequate blood flow to the pump without sucking. This newly designed inflow cannula with the beak shaped titanium tip was successfully implanted for 90 days. SN - 1058-2916 UR - https://www.unboundmedicine.com/medline/citation/15307553/Development_of_a_flexible_inflow_cannula_with_titanium_inflow_tip_for_the_NEDO_biventricular_assist_device_ DB - PRIME DP - Unbound Medicine ER -