Randomised controlled cross-over comparison of continuous positive airway pressure through the Hamilton Galileo ventilator with a Dräger CF 800 device.
Anaesthesia. 2005 Jan; 60(1):72-6.A

Abstract

In this controlled, randomised cross-over trial on 26 intensive care patients, we compared the effects on haemodynamic and respiratory profiles of continuous positive airway pressure delivered through the Hamilton Galileo ventilator or a Drager CF 800 device. We also compared the nursing time saved using the two approaches when weaning patients from mechanical ventilation. We did not find significant differences in haemodynamics, respiratory rate, physiological dead space, oxygen saturation and carbon dioxide production between the continuous positive airway pressure generated by the Galileo and Drager machines. However, there was a 10-fold reduction in nursing time using the Galileo ventilator compared with the Drager generator. We conclude that continuous positive airway pressure delivered through the Galileo ventilator is as efficient as a Drager device but consumes less nursing time.

Authors+Show Affiliations

Intensive Care Unit, Birmingham Heartlands Hospital, Birmingham Heartlands and Solihull NHS Trust (Teaching), Bordesley Green East, Birmingham, B9 5SS, England.No affiliation info availableNo affiliation info availableNo affiliation info availableNo affiliation info available

Pub Type(s)

Clinical Trial
Comparative Study
Journal Article
Randomized Controlled Trial
Research Support, Non-U.S. Gov't

Language

eng

PubMed ID

15601276

Citation

Sutton, P J., et al. "Randomised Controlled Cross-over Comparison of Continuous Positive Airway Pressure Through the Hamilton Galileo Ventilator With a Dräger CF 800 Device." Anaesthesia, vol. 60, no. 1, 2005, pp. 72-6.
Sutton PJ, Perkins CL, Giles SP, et al. Randomised controlled cross-over comparison of continuous positive airway pressure through the Hamilton Galileo ventilator with a Dräger CF 800 device. Anaesthesia. 2005;60(1):72-6.
Sutton, P. J., Perkins, C. L., Giles, S. P., McAuley, D. F., & Gao, F. (2005). Randomised controlled cross-over comparison of continuous positive airway pressure through the Hamilton Galileo ventilator with a Dräger CF 800 device. Anaesthesia, 60(1), 72-6.
Sutton PJ, et al. Randomised Controlled Cross-over Comparison of Continuous Positive Airway Pressure Through the Hamilton Galileo Ventilator With a Dräger CF 800 Device. Anaesthesia. 2005;60(1):72-6. PubMed PMID: 15601276.
* Article titles in AMA citation format should be in sentence-case
TY - JOUR T1 - Randomised controlled cross-over comparison of continuous positive airway pressure through the Hamilton Galileo ventilator with a Dräger CF 800 device. AU - Sutton,P J, AU - Perkins,C L, AU - Giles,S P, AU - McAuley,D F, AU - Gao,F, PY - 2004/12/17/pubmed PY - 2005/1/28/medline PY - 2004/12/17/entrez SP - 72 EP - 6 JF - Anaesthesia JO - Anaesthesia VL - 60 IS - 1 N2 - In this controlled, randomised cross-over trial on 26 intensive care patients, we compared the effects on haemodynamic and respiratory profiles of continuous positive airway pressure delivered through the Hamilton Galileo ventilator or a Drager CF 800 device. We also compared the nursing time saved using the two approaches when weaning patients from mechanical ventilation. We did not find significant differences in haemodynamics, respiratory rate, physiological dead space, oxygen saturation and carbon dioxide production between the continuous positive airway pressure generated by the Galileo and Drager machines. However, there was a 10-fold reduction in nursing time using the Galileo ventilator compared with the Drager generator. We conclude that continuous positive airway pressure delivered through the Galileo ventilator is as efficient as a Drager device but consumes less nursing time. SN - 0003-2409 UR - https://www.unboundmedicine.com/medline/citation/15601276/Randomised_controlled_cross_over_comparison_of_continuous_positive_airway_pressure_through_the_Hamilton_Galileo_ventilator_with_a_Dr��ger_CF_800_device_ DB - PRIME DP - Unbound Medicine ER -