Cadaver fitting study of the DexAide right ventricular assist device.
Artif Organs. 2007 Aug; 31(8):646-8.AO

Abstract

The DexAide right ventricular assist device (RVAD) has been developed to provide an implantable RVAD option to surgeons. The aim of this study was to determine the optimal cannula design and optimal implantation location of the DexAide RVAD in preparation for its clinical use. Separately, a HeartMate XVE left ventricular assist device (LVAD) and CorAide LVAD models were implanted into the preperitoneal and right thoracic space, and the anatomical fit of the DexAide RVAD was evaluated in five preserved human cadavers. The DexAide RVAD inflow cannula was inserted through the diaphragmatic surface of the right ventricle and the outflow was directed to the pulmonary artery. Right thoracic implantation of the DexAide RVAD provided an excellent fit with either the HeartMate or CorAide LVAD in all cadavers. The results of this study will guide improvements in the designs of cannulae and implantation of the DexAide RVAD in future clinical applications.

Authors+Show Affiliations

Department of Biomedical Engineering, Lerner Research Institute, Cleveland Clinic, Cleveland, OH, USA.No affiliation info availableNo affiliation info availableNo affiliation info availableNo affiliation info availableNo affiliation info availableNo affiliation info availableNo affiliation info availableNo affiliation info availableNo affiliation info availableNo affiliation info availableNo affiliation info availableNo affiliation info availableNo affiliation info availableNo affiliation info available

Pub Type(s)

Journal Article

Language

eng

PubMed ID

17651120

Citation

Ootaki, Yoshio, et al. "Cadaver Fitting Study of the DexAide Right Ventricular Assist Device." Artificial Organs, vol. 31, no. 8, 2007, pp. 646-8.
Ootaki Y, Ootaki C, Kamohara K, et al. Cadaver fitting study of the DexAide right ventricular assist device. Artif Organs. 2007;31(8):646-8.
Ootaki, Y., Ootaki, C., Kamohara, K., Akiyama, M., Cingoz, F., Kopcak, M. W., Dessoffy, R., Liu, J., Massiello, A., Horvath, D., Chen, J. F., Zhou, Q., Benefit, S., Golding, L. A., & Fukamachi, K. (2007). Cadaver fitting study of the DexAide right ventricular assist device. Artificial Organs, 31(8), 646-8.
Ootaki Y, et al. Cadaver Fitting Study of the DexAide Right Ventricular Assist Device. Artif Organs. 2007;31(8):646-8. PubMed PMID: 17651120.
* Article titles in AMA citation format should be in sentence-case
TY - JOUR T1 - Cadaver fitting study of the DexAide right ventricular assist device. AU - Ootaki,Yoshio, AU - Ootaki,Chiyo, AU - Kamohara,Keiji, AU - Akiyama,Masatoshi, AU - Cingoz,Faruk, AU - Kopcak,Michael W,Jr AU - Dessoffy,Raymond, AU - Liu,Jenny, AU - Massiello,Alex, AU - Horvath,David, AU - Chen,Ji-Feng, AU - Zhou,Qun, AU - Benefit,Stephen, AU - Golding,Leonard A R, AU - Fukamachi,Kiyotaka, PY - 2007/7/27/pubmed PY - 2007/10/3/medline PY - 2007/7/27/entrez SP - 646 EP - 8 JF - Artificial organs JO - Artif Organs VL - 31 IS - 8 N2 - The DexAide right ventricular assist device (RVAD) has been developed to provide an implantable RVAD option to surgeons. The aim of this study was to determine the optimal cannula design and optimal implantation location of the DexAide RVAD in preparation for its clinical use. Separately, a HeartMate XVE left ventricular assist device (LVAD) and CorAide LVAD models were implanted into the preperitoneal and right thoracic space, and the anatomical fit of the DexAide RVAD was evaluated in five preserved human cadavers. The DexAide RVAD inflow cannula was inserted through the diaphragmatic surface of the right ventricle and the outflow was directed to the pulmonary artery. Right thoracic implantation of the DexAide RVAD provided an excellent fit with either the HeartMate or CorAide LVAD in all cadavers. The results of this study will guide improvements in the designs of cannulae and implantation of the DexAide RVAD in future clinical applications. SN - 0160-564X UR - https://www.unboundmedicine.com/medline/citation/17651120/Cadaver_fitting_study_of_the_DexAide_right_ventricular_assist_device_ L2 - https://doi.org/10.1111/j.1525-1594.2007.00438.x DB - PRIME DP - Unbound Medicine ER -