Comparison of dynamic contour tonometry with Goldmann applanation tonometry in glaucoma practice.
Acta Ophthalmol 2009; 87(3):323-8AO

Abstract

PURPOSE

To compare intraocular pressure (IOP) readings taken using dynamic contour tonometry (DCT) with IOP readings taken with Goldmann applanation tonometry (GAT) in eyes with glaucoma or ocular hypertension.

METHODS

The present study included 100 eyes in 100 patients with glaucoma or ocular hypertension. After pachymetry DCT and GAT were performed. Intraocular pressures as measured with DCT and GAT were compared with one another and with central corneal thickness (CCT).

RESULTS

Mean DCT IOP measurements (20.1 +/- 4.3 mmHg) were significantly (p < 0.001) higher than GAT IOP values (17.9 +/- 4.7 mmHg). The mean difference between DCT and GAT measurements was 2.1 mmHg (range -3.4 to 9.7 mmHg). The difference followed a normal distribution. Measurements made with DCT and GAT correlated significantly with one another (Spearman's rho = 0.761, p < 0.001). Neither GAT nor DCT measurements showed a significant correlation with CCT (537 +/- 39 microm, range 458-656 microm). Multivariate regression analysis has shown that the difference between DCT and GAT is influenced significantly by ocular pulse amplitude (r = -0.334, p = 0.001) and it is not influenced by CCT (r = -0.106, p = 0.292).

CONCLUSIONS

In eyes with glaucoma or ocular hypertension, DCT facilitates suitable and reliable IOP measurements which are in good concordance with GAT readings. Variation in CCT cannot by itself explain the differences in measurements taken with DCT and GAT in a number of eyes.

Authors+Show Affiliations

Ophthalmiatrion Athinon (Athens Eye Hospital), Athens, Greece. ihalkia@otenet.grNo affiliation info availableNo affiliation info availableNo affiliation info availableNo affiliation info availableNo affiliation info availableNo affiliation info availableNo affiliation info availableNo affiliation info availableNo affiliation info available

Pub Type(s)

Comparative Study
Evaluation Studies
Journal Article

Language

eng

PubMed ID

18631335

Citation

Halkiadakis, Ioannis, et al. "Comparison of Dynamic Contour Tonometry With Goldmann Applanation Tonometry in Glaucoma Practice." Acta Ophthalmologica, vol. 87, no. 3, 2009, pp. 323-8.
Halkiadakis I, Patsea E, Chatzimichali K, et al. Comparison of dynamic contour tonometry with Goldmann applanation tonometry in glaucoma practice. Acta Ophthalmol. 2009;87(3):323-8.
Halkiadakis, I., Patsea, E., Chatzimichali, K., Skouriotis, S., Chalkidou, S., Amariotakis, G., ... Georgopoulos, G. (2009). Comparison of dynamic contour tonometry with Goldmann applanation tonometry in glaucoma practice. Acta Ophthalmologica, 87(3), pp. 323-8. doi:10.1111/j.1755-3768.2008.01239.x.
Halkiadakis I, et al. Comparison of Dynamic Contour Tonometry With Goldmann Applanation Tonometry in Glaucoma Practice. Acta Ophthalmol. 2009;87(3):323-8. PubMed PMID: 18631335.
* Article titles in AMA citation format should be in sentence-case
TY - JOUR T1 - Comparison of dynamic contour tonometry with Goldmann applanation tonometry in glaucoma practice. AU - Halkiadakis,Ioannis, AU - Patsea,Eleni, AU - Chatzimichali,Katerina, AU - Skouriotis,Sotiris, AU - Chalkidou,Sonia, AU - Amariotakis,Georgios, AU - Papakonstadinou,Dimitrios, AU - Theodossiadis,George, AU - Amariotakis,Apostolos, AU - Georgopoulos,Gerasimos, Y1 - 2008/07/09/ PY - 2008/7/18/pubmed PY - 2009/7/29/medline PY - 2008/7/18/entrez SP - 323 EP - 8 JF - Acta ophthalmologica JO - Acta Ophthalmol VL - 87 IS - 3 N2 - PURPOSE: To compare intraocular pressure (IOP) readings taken using dynamic contour tonometry (DCT) with IOP readings taken with Goldmann applanation tonometry (GAT) in eyes with glaucoma or ocular hypertension. METHODS: The present study included 100 eyes in 100 patients with glaucoma or ocular hypertension. After pachymetry DCT and GAT were performed. Intraocular pressures as measured with DCT and GAT were compared with one another and with central corneal thickness (CCT). RESULTS: Mean DCT IOP measurements (20.1 +/- 4.3 mmHg) were significantly (p < 0.001) higher than GAT IOP values (17.9 +/- 4.7 mmHg). The mean difference between DCT and GAT measurements was 2.1 mmHg (range -3.4 to 9.7 mmHg). The difference followed a normal distribution. Measurements made with DCT and GAT correlated significantly with one another (Spearman's rho = 0.761, p < 0.001). Neither GAT nor DCT measurements showed a significant correlation with CCT (537 +/- 39 microm, range 458-656 microm). Multivariate regression analysis has shown that the difference between DCT and GAT is influenced significantly by ocular pulse amplitude (r = -0.334, p = 0.001) and it is not influenced by CCT (r = -0.106, p = 0.292). CONCLUSIONS: In eyes with glaucoma or ocular hypertension, DCT facilitates suitable and reliable IOP measurements which are in good concordance with GAT readings. Variation in CCT cannot by itself explain the differences in measurements taken with DCT and GAT in a number of eyes. SN - 1755-3768 UR - https://www.unboundmedicine.com/medline/citation/18631335/Comparison_of_dynamic_contour_tonometry_with_Goldmann_applanation_tonometry_in_glaucoma_practice_ L2 - https://doi.org/10.1111/j.1755-3768.2008.01239.x DB - PRIME DP - Unbound Medicine ER -