Anterior segment imaging: Fourier-domain optical coherence tomography versus time-domain optical coherence tomography.
J Cataract Refract Surg. 2009 Aug; 35(8):1410-4.JC

Abstract

PURPOSE

To compare anterior segment measurements and morphology of 2 optical coherence tomography (OCT) systems.

SETTING

Department of Ophthalmology, District Railway Hospital, and the Nursing Department and Social Medical Issues, Health Care Division, Silesian Medical University, Katowice, Poland.

METHODS

In normal eyes and in eyes with corneal and trabecular-iris angle disorders, the central corneal thickness (CCT), trabecular-iris angle, and angle-opening distance at the nasal and temporal angles were measured 3 times during 1 visit using the Visante time-domain OCT system and the RTVue-100 Fourier-domain corneal anterior module OCT system. Anterior segment morphology was assessed and compared.

RESULTS

Fifty-four eyes were evaluated. The mean values (+/-SD) by time-domain OCT and Fourier-domain OCT were, respectively, automatic CCT, 535 +/- 33.07 microm and 538 +/- 31.82 microm; manual CCT, 545 +/- 30.91 microm and 542 +/- 30.57 microm; nasal trabecular-iris angle, 34.7 +/- 9.5 degrees and 35.2 +/- 8.9 degrees; temporal trabecular-iris angle, 35.3 +/- 8.5 degrees and 35.5 +/- 9 degrees; nasal angle-opening distance, 435 +/- 95 microm and 444 +/- 98 microm; and temporal angle-opening distance, 443 +/- 103 microm and 452 +/- 99 microm. There was no significant difference between mean values, and they were highly correlated. On morphologic analysis, time-domain OCT had lower resolution; however, all anterior chamber structures were visible on 1 image. Fourier-domain OCT provided precise information about small areas of the anterior chamber.

CONCLUSION

Fourier-domain OCT provided accurate anterior eye segment measurements that agreed with those obtained with time-domain OCT.

Authors+Show Affiliations

Department of Ophthalmology, District Railway Hospital, Silesian Medical University, Katowice, Poland.No affiliation info availableNo affiliation info availableNo affiliation info availableNo affiliation info available

Pub Type(s)

Comparative Study
Journal Article

Language

eng

PubMed ID

19631129

Citation

Wylegała, Edward, et al. "Anterior Segment Imaging: Fourier-domain Optical Coherence Tomography Versus Time-domain Optical Coherence Tomography." Journal of Cataract and Refractive Surgery, vol. 35, no. 8, 2009, pp. 1410-4.
Wylegała E, Teper S, Nowińska AK, et al. Anterior segment imaging: Fourier-domain optical coherence tomography versus time-domain optical coherence tomography. J Cataract Refract Surg. 2009;35(8):1410-4.
Wylegała, E., Teper, S., Nowińska, A. K., Milka, M., & Dobrowolski, D. (2009). Anterior segment imaging: Fourier-domain optical coherence tomography versus time-domain optical coherence tomography. Journal of Cataract and Refractive Surgery, 35(8), 1410-4. https://doi.org/10.1016/j.jcrs.2009.03.034
Wylegała E, et al. Anterior Segment Imaging: Fourier-domain Optical Coherence Tomography Versus Time-domain Optical Coherence Tomography. J Cataract Refract Surg. 2009;35(8):1410-4. PubMed PMID: 19631129.
* Article titles in AMA citation format should be in sentence-case
TY - JOUR T1 - Anterior segment imaging: Fourier-domain optical coherence tomography versus time-domain optical coherence tomography. AU - Wylegała,Edward, AU - Teper,Slawomir, AU - Nowińska,Anna K, AU - Milka,Michal, AU - Dobrowolski,Dariusz, PY - 2008/11/02/received PY - 2009/03/19/revised PY - 2009/03/20/accepted PY - 2009/7/28/entrez PY - 2009/7/28/pubmed PY - 2009/10/13/medline SP - 1410 EP - 4 JF - Journal of cataract and refractive surgery JO - J Cataract Refract Surg VL - 35 IS - 8 N2 - PURPOSE: To compare anterior segment measurements and morphology of 2 optical coherence tomography (OCT) systems. SETTING: Department of Ophthalmology, District Railway Hospital, and the Nursing Department and Social Medical Issues, Health Care Division, Silesian Medical University, Katowice, Poland. METHODS: In normal eyes and in eyes with corneal and trabecular-iris angle disorders, the central corneal thickness (CCT), trabecular-iris angle, and angle-opening distance at the nasal and temporal angles were measured 3 times during 1 visit using the Visante time-domain OCT system and the RTVue-100 Fourier-domain corneal anterior module OCT system. Anterior segment morphology was assessed and compared. RESULTS: Fifty-four eyes were evaluated. The mean values (+/-SD) by time-domain OCT and Fourier-domain OCT were, respectively, automatic CCT, 535 +/- 33.07 microm and 538 +/- 31.82 microm; manual CCT, 545 +/- 30.91 microm and 542 +/- 30.57 microm; nasal trabecular-iris angle, 34.7 +/- 9.5 degrees and 35.2 +/- 8.9 degrees; temporal trabecular-iris angle, 35.3 +/- 8.5 degrees and 35.5 +/- 9 degrees; nasal angle-opening distance, 435 +/- 95 microm and 444 +/- 98 microm; and temporal angle-opening distance, 443 +/- 103 microm and 452 +/- 99 microm. There was no significant difference between mean values, and they were highly correlated. On morphologic analysis, time-domain OCT had lower resolution; however, all anterior chamber structures were visible on 1 image. Fourier-domain OCT provided precise information about small areas of the anterior chamber. CONCLUSION: Fourier-domain OCT provided accurate anterior eye segment measurements that agreed with those obtained with time-domain OCT. SN - 1873-4502 UR - https://www.unboundmedicine.com/medline/citation/19631129/Anterior_segment_imaging:_Fourier_domain_optical_coherence_tomography_versus_time_domain_optical_coherence_tomography_ L2 - https://linkinghub.elsevier.com/retrieve/pii/S0886-3350(09)00479-9 DB - PRIME DP - Unbound Medicine ER -