Novel contact force sensor incorporated in irrigated radiofrequency ablation catheter predicts lesion size and incidence of steam pop and thrombus.
Circ Arrhythm Electrophysiol. 2008 Dec; 1(5):354-62.CA

Abstract

BACKGROUND

An open-irrigated radiofrequency (RF) ablation catheter was developed to measure contact force (CF). Three optical fibers measure microdeformation of the catheter tip. The purpose of this study was to (1) validate the accuracy of CF sensor (CFS) (bench test); and (2) determine the relationship between CF and tissue temperatures, lesion size, steam pop, and thrombus during RF ablation using a canine thigh muscle preparation.

METHODS AND RESULTS

CFS measurements (total 1409) from 2 catheters in 3 angles (perpendicular, parallel, and 45 degrees) were compared with a certified balance (range, 0 to 50 g). CFS measurements correlated highly (R(2) > or =0.988; mean error, < or =1.0 g). In 10 anesthetized dogs, a skin cradle over the thigh muscle was superfused with heparinized blood at 37 degrees C. A 7F catheter with 3.5-mm saline-irrigated electrode and CFS (Endosense) was held perpendicular to the muscle at CF of 2, 10, 20, 30, and 40 g. RF was delivered (n=100) for 60 seconds at 30 or 50 W (irrigation 17 or 30 mL/min). Tissue temperature (3 and 7 mm depths), lesion size, thrombus, and steam pop increased significantly with increasing CF at each RF power. Lesion size was greater with applications of lower power (30 W) and greater CF (30 to 40 g) than at high power (50 W) with lower CF (2 to 10 g).

CONCLUSIONS

This novel ablation catheter, which accurately measures CF, confirmed CF is a major determinant of RF lesion size. Steam pop and thrombus incidence also increases with CF. CFS in an open-irrigated ablation catheter that may optimize the selection of RF power and application time to maximize lesion formation and reduce the risk of steam pop and thrombus.

Authors+Show Affiliations

Heart Rhythm Institute and Department of Medicine, University of Oklahoma Health Sciences Center, Oklahoma City, Oklahoma 73104, USA.No affiliation info availableNo affiliation info availableNo affiliation info availableNo affiliation info availableNo affiliation info availableNo affiliation info availableNo affiliation info availableNo affiliation info availableNo affiliation info availableNo affiliation info available

Pub Type(s)

Journal Article
Research Support, Non-U.S. Gov't
Validation Study

Language

eng

PubMed ID

19808430

Citation

Yokoyama, Katsuaki, et al. "Novel Contact Force Sensor Incorporated in Irrigated Radiofrequency Ablation Catheter Predicts Lesion Size and Incidence of Steam Pop and Thrombus." Circulation. Arrhythmia and Electrophysiology, vol. 1, no. 5, 2008, pp. 354-62.
Yokoyama K, Nakagawa H, Shah DC, et al. Novel contact force sensor incorporated in irrigated radiofrequency ablation catheter predicts lesion size and incidence of steam pop and thrombus. Circ Arrhythm Electrophysiol. 2008;1(5):354-62.
Yokoyama, K., Nakagawa, H., Shah, D. C., Lambert, H., Leo, G., Aeby, N., Ikeda, A., Pitha, J. V., Sharma, T., Lazzara, R., & Jackman, W. M. (2008). Novel contact force sensor incorporated in irrigated radiofrequency ablation catheter predicts lesion size and incidence of steam pop and thrombus. Circulation. Arrhythmia and Electrophysiology, 1(5), 354-62. https://doi.org/10.1161/CIRCEP.108.803650
Yokoyama K, et al. Novel Contact Force Sensor Incorporated in Irrigated Radiofrequency Ablation Catheter Predicts Lesion Size and Incidence of Steam Pop and Thrombus. Circ Arrhythm Electrophysiol. 2008;1(5):354-62. PubMed PMID: 19808430.
* Article titles in AMA citation format should be in sentence-case
TY - JOUR T1 - Novel contact force sensor incorporated in irrigated radiofrequency ablation catheter predicts lesion size and incidence of steam pop and thrombus. AU - Yokoyama,Katsuaki, AU - Nakagawa,Hiroshi, AU - Shah,Dipen C, AU - Lambert,Hendrik, AU - Leo,Giovanni, AU - Aeby,Nicolas, AU - Ikeda,Atsushi, AU - Pitha,Jan V, AU - Sharma,Tushar, AU - Lazzara,Ralph, AU - Jackman,Warren M, Y1 - 2008/12/02/ PY - 2009/10/8/entrez PY - 2009/10/8/pubmed PY - 2009/10/29/medline SP - 354 EP - 62 JF - Circulation. Arrhythmia and electrophysiology JO - Circ Arrhythm Electrophysiol VL - 1 IS - 5 N2 - BACKGROUND: An open-irrigated radiofrequency (RF) ablation catheter was developed to measure contact force (CF). Three optical fibers measure microdeformation of the catheter tip. The purpose of this study was to (1) validate the accuracy of CF sensor (CFS) (bench test); and (2) determine the relationship between CF and tissue temperatures, lesion size, steam pop, and thrombus during RF ablation using a canine thigh muscle preparation. METHODS AND RESULTS: CFS measurements (total 1409) from 2 catheters in 3 angles (perpendicular, parallel, and 45 degrees) were compared with a certified balance (range, 0 to 50 g). CFS measurements correlated highly (R(2) > or =0.988; mean error, < or =1.0 g). In 10 anesthetized dogs, a skin cradle over the thigh muscle was superfused with heparinized blood at 37 degrees C. A 7F catheter with 3.5-mm saline-irrigated electrode and CFS (Endosense) was held perpendicular to the muscle at CF of 2, 10, 20, 30, and 40 g. RF was delivered (n=100) for 60 seconds at 30 or 50 W (irrigation 17 or 30 mL/min). Tissue temperature (3 and 7 mm depths), lesion size, thrombus, and steam pop increased significantly with increasing CF at each RF power. Lesion size was greater with applications of lower power (30 W) and greater CF (30 to 40 g) than at high power (50 W) with lower CF (2 to 10 g). CONCLUSIONS: This novel ablation catheter, which accurately measures CF, confirmed CF is a major determinant of RF lesion size. Steam pop and thrombus incidence also increases with CF. CFS in an open-irrigated ablation catheter that may optimize the selection of RF power and application time to maximize lesion formation and reduce the risk of steam pop and thrombus. SN - 1941-3084 UR - https://www.unboundmedicine.com/medline/citation/19808430/Novel_contact_force_sensor_incorporated_in_irrigated_radiofrequency_ablation_catheter_predicts_lesion_size_and_incidence_of_steam_pop_and_thrombus_ L2 - https://www.ahajournals.org/doi/10.1161/CIRCEP.108.803650?url_ver=Z39.88-2003&amp;rfr_id=ori:rid:crossref.org&amp;rfr_dat=cr_pub=pubmed DB - PRIME DP - Unbound Medicine ER -