Mechanical effects of implant-tooth rigid connection by a fixed partial denture: a 3d finite element analysis.Eur J Prosthodont Restor Dent. 2009 Sep; 17(3):98-104.EJ
Abstract
This study investigated the mechanical behaviour of tooth-implant fixed partial dentures by three-dimensional Finite Element Analysis. Titanium implants of various length and diameter were rigidly connected to a premolar either by two connected ceramic crowns or by a three unit Fixed Partial Denture. The implant acted like the fulcrum of an orthodontic device inducing tooth intrusion. Bone stresses appeared around the neck of the implant. Implant and abutment's stresses were meaningfully less intense with high diameter and low span. Implant length was a less influencing parameter. The use of wide-bodied implants is recommended in selected cases of short-span fixed partial dentures.
MeSH
Pub Type(s)
Journal Article
Language
eng
PubMed ID
19839184
Citation
Naveau, Adrien, and Laurent Pierrisnard. "Mechanical Effects of Implant-tooth Rigid Connection By a Fixed Partial Denture: a 3d Finite Element Analysis." The European Journal of Prosthodontics and Restorative Dentistry, vol. 17, no. 3, 2009, pp. 98-104.
Naveau A, Pierrisnard L. Mechanical effects of implant-tooth rigid connection by a fixed partial denture: a 3d finite element analysis. Eur J Prosthodont Restor Dent. 2009;17(3):98-104.
Naveau, A., & Pierrisnard, L. (2009). Mechanical effects of implant-tooth rigid connection by a fixed partial denture: a 3d finite element analysis. The European Journal of Prosthodontics and Restorative Dentistry, 17(3), 98-104.
Naveau A, Pierrisnard L. Mechanical Effects of Implant-tooth Rigid Connection By a Fixed Partial Denture: a 3d Finite Element Analysis. Eur J Prosthodont Restor Dent. 2009;17(3):98-104. PubMed PMID: 19839184.
* Article titles in AMA citation format should be in sentence-case
TY - JOUR
T1 - Mechanical effects of implant-tooth rigid connection by a fixed partial denture: a 3d finite element analysis.
AU - Naveau,Adrien,
AU - Pierrisnard,Laurent,
PY - 2009/10/21/entrez
PY - 2009/10/21/pubmed
PY - 2010/1/20/medline
SP - 98
EP - 104
JF - The European journal of prosthodontics and restorative dentistry
JO - Eur J Prosthodont Restor Dent
VL - 17
IS - 3
N2 - This study investigated the mechanical behaviour of tooth-implant fixed partial dentures by three-dimensional Finite Element Analysis. Titanium implants of various length and diameter were rigidly connected to a premolar either by two connected ceramic crowns or by a three unit Fixed Partial Denture. The implant acted like the fulcrum of an orthodontic device inducing tooth intrusion. Bone stresses appeared around the neck of the implant. Implant and abutment's stresses were meaningfully less intense with high diameter and low span. Implant length was a less influencing parameter. The use of wide-bodied implants is recommended in selected cases of short-span fixed partial dentures.
SN - 0965-7452
UR - https://www.unboundmedicine.com/medline/citation/19839184/Mechanical_effects_of_implant_tooth_rigid_connection_by_a_fixed_partial_denture:_a_3d_finite_element_analysis_
DB - PRIME
DP - Unbound Medicine
ER -