Biocompatibility and mechanical properties of diamond-like coatings on cobalt-chromium-molybdenum steel and titanium-aluminum-vanadium biomedical alloys.
J Biomed Mater Res A. 2010 Nov; 95(2):388-400.JB

Abstract

Diamond-like carbon (DLC) films are favored for wear components because of diamond-like hardness, low friction, low wear, and high corrosion resistance (Schultz et al., Mat-wiss u Werkstofftech 2004;35:924-928; Lappalainen et al., J Biomed Mater Res B Appl Biomater 2003;66B:410-413; Tiainen, Diam Relat Mater 2001;10:153-160). Several studies have demonstrated their inertness, nontoxicity, and the biocompatibility, which has led to interest among manufacturers of surgical implants (Allen et al., J Biomed Mater Res B Appl Biomater 2001;58:319-328; Uzumaki et al., Diam Relat Mater 2006;15:982-988; Hauert, Diam Relat Mater 2003;12:583-589; Grill, Diam Relat Mater 2003;12:166-170). In this study, hydrogen-free amorphous, tetrahedrally bonded DLC films (ta-C) were deposited at low temperatures by physical vapor deposition on medical grade Co28Cr6Mo steel and the titanium alloy Ti6Al4V (Scheibe et al., Surf Coat Tech 1996;85:209-214). The mechanical performance of the ta-C was characterized by measuring its surface roughness, contact angle, adhesion, and wear behavior, whereas the biocompatibility was assessed by osteoblast (OB) attachment and cell viability via Live/Dead assay. There was no statistical difference found in the wettability as measured by contact angle measurements for the ta-C coated and the uncoated samples of either Co28Cr6Mo or Ti6Al4V. Rockwell C indentation and dynamic scratch testing on 2-10 μm thick ta-C films on Co28Cr6Mo substrates showed excellent adhesion with HF1 grade and up to 48 N for the critical load L(C2) during scratch testing. The ta-C coating reduced the wear from 3.5 × 10(-5) mm(3)/Nm for an uncoated control sample (uncoated Co28Cr6Mo against uncoated stainless steel) to 1.1 × 10(-7) mm(3)/Nm (coated Co28Cr6Mo against uncoated stainless steel) in reciprocating pin-on-disk testing. The lowest wear factor of 3.9 × 10(-10) mm(3)/Nm was measured using a ta-C coated steel ball running against a ta-C coated and polished Co28Cr6Mo disk. Student's t-test found that the ta-C coating had no statistically significant (p < 0.05) effect on OB attachment, when compared with the uncoated control samples. There was no significant difference (p < 0.05) in the Live/Dead assay results in cell death between the ta-C coated Co28Cr6Mo and Ti6Al4V samples and the uncoated controls. Therefore, these ta-C coatings show improved wear and corrosion (Dorner-Reisel et al., Diam Relat Mater 2003;11:823-827; Affato et al., J Biomed Mater Res B Appl Biomater 2000;53:221-226; Dorner-Reisel et al., Surf Coat Tech 2004;177-178:830-837; Kim et al., Diam Relat Mater 2004;14:35-41) performance and excellent in vitro cyto-compatibility, when compared with currently used uncoated Co28Cr6Mo and Ti6Al4V implant materials.

Authors+Show Affiliations

Fraunhofer Center for Coatings and Laser Applications, East Lansing, Michigan 48823, USA.No affiliation info availableNo affiliation info availableNo affiliation info availableNo affiliation info availableNo affiliation info availableNo affiliation info availableNo affiliation info available

Pub Type(s)

Evaluation Study
Journal Article
Research Support, Non-U.S. Gov't

Language

eng

PubMed ID

20648536

Citation

Hinüber, C, et al. "Biocompatibility and Mechanical Properties of Diamond-like Coatings On Cobalt-chromium-molybdenum Steel and Titanium-aluminum-vanadium Biomedical Alloys." Journal of Biomedical Materials Research. Part A, vol. 95, no. 2, 2010, pp. 388-400.
Hinüber C, Kleemann C, Friederichs RJ, et al. Biocompatibility and mechanical properties of diamond-like coatings on cobalt-chromium-molybdenum steel and titanium-aluminum-vanadium biomedical alloys. J Biomed Mater Res A. 2010;95(2):388-400.
Hinüber, C., Kleemann, C., Friederichs, R. J., Haubold, L., Scheibe, H. J., Schuelke, T., Boehlert, C., & Baumann, M. J. (2010). Biocompatibility and mechanical properties of diamond-like coatings on cobalt-chromium-molybdenum steel and titanium-aluminum-vanadium biomedical alloys. Journal of Biomedical Materials Research. Part A, 95(2), 388-400. https://doi.org/10.1002/jbm.a.32851
Hinüber C, et al. Biocompatibility and Mechanical Properties of Diamond-like Coatings On Cobalt-chromium-molybdenum Steel and Titanium-aluminum-vanadium Biomedical Alloys. J Biomed Mater Res A. 2010;95(2):388-400. PubMed PMID: 20648536.
* Article titles in AMA citation format should be in sentence-case
TY - JOUR T1 - Biocompatibility and mechanical properties of diamond-like coatings on cobalt-chromium-molybdenum steel and titanium-aluminum-vanadium biomedical alloys. AU - Hinüber,C, AU - Kleemann,C, AU - Friederichs,R J, AU - Haubold,L, AU - Scheibe,H J, AU - Schuelke,T, AU - Boehlert,C, AU - Baumann,M J, PY - 2010/7/22/entrez PY - 2010/7/22/pubmed PY - 2011/1/14/medline SP - 388 EP - 400 JF - Journal of biomedical materials research. Part A JO - J Biomed Mater Res A VL - 95 IS - 2 N2 - Diamond-like carbon (DLC) films are favored for wear components because of diamond-like hardness, low friction, low wear, and high corrosion resistance (Schultz et al., Mat-wiss u Werkstofftech 2004;35:924-928; Lappalainen et al., J Biomed Mater Res B Appl Biomater 2003;66B:410-413; Tiainen, Diam Relat Mater 2001;10:153-160). Several studies have demonstrated their inertness, nontoxicity, and the biocompatibility, which has led to interest among manufacturers of surgical implants (Allen et al., J Biomed Mater Res B Appl Biomater 2001;58:319-328; Uzumaki et al., Diam Relat Mater 2006;15:982-988; Hauert, Diam Relat Mater 2003;12:583-589; Grill, Diam Relat Mater 2003;12:166-170). In this study, hydrogen-free amorphous, tetrahedrally bonded DLC films (ta-C) were deposited at low temperatures by physical vapor deposition on medical grade Co28Cr6Mo steel and the titanium alloy Ti6Al4V (Scheibe et al., Surf Coat Tech 1996;85:209-214). The mechanical performance of the ta-C was characterized by measuring its surface roughness, contact angle, adhesion, and wear behavior, whereas the biocompatibility was assessed by osteoblast (OB) attachment and cell viability via Live/Dead assay. There was no statistical difference found in the wettability as measured by contact angle measurements for the ta-C coated and the uncoated samples of either Co28Cr6Mo or Ti6Al4V. Rockwell C indentation and dynamic scratch testing on 2-10 μm thick ta-C films on Co28Cr6Mo substrates showed excellent adhesion with HF1 grade and up to 48 N for the critical load L(C2) during scratch testing. The ta-C coating reduced the wear from 3.5 × 10(-5) mm(3)/Nm for an uncoated control sample (uncoated Co28Cr6Mo against uncoated stainless steel) to 1.1 × 10(-7) mm(3)/Nm (coated Co28Cr6Mo against uncoated stainless steel) in reciprocating pin-on-disk testing. The lowest wear factor of 3.9 × 10(-10) mm(3)/Nm was measured using a ta-C coated steel ball running against a ta-C coated and polished Co28Cr6Mo disk. Student's t-test found that the ta-C coating had no statistically significant (p < 0.05) effect on OB attachment, when compared with the uncoated control samples. There was no significant difference (p < 0.05) in the Live/Dead assay results in cell death between the ta-C coated Co28Cr6Mo and Ti6Al4V samples and the uncoated controls. Therefore, these ta-C coatings show improved wear and corrosion (Dorner-Reisel et al., Diam Relat Mater 2003;11:823-827; Affato et al., J Biomed Mater Res B Appl Biomater 2000;53:221-226; Dorner-Reisel et al., Surf Coat Tech 2004;177-178:830-837; Kim et al., Diam Relat Mater 2004;14:35-41) performance and excellent in vitro cyto-compatibility, when compared with currently used uncoated Co28Cr6Mo and Ti6Al4V implant materials. SN - 1552-4965 UR - https://www.unboundmedicine.com/medline/citation/20648536/Biocompatibility_and_mechanical_properties_of_diamond_like_coatings_on_cobalt_chromium_molybdenum_steel_and_titanium_aluminum_vanadium_biomedical_alloys_ L2 - https://doi.org/10.1002/jbm.a.32851 DB - PRIME DP - Unbound Medicine ER -