Technique for insertion of HeartMate II left ventricular assist device inflow cannula.Ann Thorac Surg. 2011 Jun; 91(6):2001-2.AT
Abstract
During implantation of the HeartMate II left ventricular assist device, the inflow cannula is typically inserted in such a way that the sewing ring is placed within the ventricular cavity in contact with the endocardium. We describe a novel implantation technique in which the sewing ring is secured to the left ventricular epicardium, thus avoiding the complication of bleeding from the exposed cut surface of apical myocardium. In addition, the shortened intraventricular segment of the inflow cannula protects against inflow occlusion by the interventricular septum.
Links
Pub Type(s)
Journal Article
Language
eng
PubMed ID
21620010
Citation
Soleimani, Behzad, et al. "Technique for Insertion of HeartMate II Left Ventricular Assist Device Inflow Cannula." The Annals of Thoracic Surgery, vol. 91, no. 6, 2011, pp. 2001-2.
Soleimani B, Stephenson ER, Pae WE. Technique for insertion of HeartMate II left ventricular assist device inflow cannula. Ann Thorac Surg. 2011;91(6):2001-2.
Soleimani, B., Stephenson, E. R., & Pae, W. E. (2011). Technique for insertion of HeartMate II left ventricular assist device inflow cannula. The Annals of Thoracic Surgery, 91(6), 2001-2. https://doi.org/10.1016/j.athoracsur.2010.12.059
Soleimani B, Stephenson ER, Pae WE. Technique for Insertion of HeartMate II Left Ventricular Assist Device Inflow Cannula. Ann Thorac Surg. 2011;91(6):2001-2. PubMed PMID: 21620010.
* Article titles in AMA citation format should be in sentence-case
TY - JOUR
T1 - Technique for insertion of HeartMate II left ventricular assist device inflow cannula.
AU - Soleimani,Behzad,
AU - Stephenson,Edward R,
AU - Pae,Walter E,
PY - 2010/11/29/received
PY - 2010/11/29/revised
PY - 2010/12/30/accepted
PY - 2011/5/31/entrez
PY - 2011/5/31/pubmed
PY - 2011/8/13/medline
SP - 2001
EP - 2
JF - The Annals of thoracic surgery
JO - Ann Thorac Surg
VL - 91
IS - 6
N2 - During implantation of the HeartMate II left ventricular assist device, the inflow cannula is typically inserted in such a way that the sewing ring is placed within the ventricular cavity in contact with the endocardium. We describe a novel implantation technique in which the sewing ring is secured to the left ventricular epicardium, thus avoiding the complication of bleeding from the exposed cut surface of apical myocardium. In addition, the shortened intraventricular segment of the inflow cannula protects against inflow occlusion by the interventricular septum.
SN - 1552-6259
UR - https://www.unboundmedicine.com/medline/citation/21620010/Technique_for_insertion_of_HeartMate_II_left_ventricular_assist_device_inflow_cannula_
L2 - https://linkinghub.elsevier.com/retrieve/pii/S0003-4975(11)00012-9
DB - PRIME
DP - Unbound Medicine
ER -