Measurement of the Optic Nerve Sheath Diameter in Children: Comparison Between Transbulbar Sonography and Magnetic Resonance Imaging.
Ultraschall Med. 2012 Dec; 33(6):569-573.UM

Abstract

PURPOSE

To evaluate the accuracy of measurements of the optic nerve sheath diameter (ONSD) in children by comparing transbulbar sonography with magnetic resonance imaging (MRI).

MATERIALS AND METHODS

65 children (age: 3 mo- 17y; mean age: 11.3y) underwent MR imaging of the brain including a heavily T 2-w sequence of the orbit and transbulbar sonography using a 17 MHz linear array transducer. Measurements of the ONSD were performed retrospectively by two experienced readers and all images were evaluated regarding the image quality. Bland-Altman Plots were produced to assess the accuracy of measurements. The correlation between readers and between MR imaging and transbulbar sonography was calculated by the concordance correlation coefficient (CCC).

RESULTS

Overall the mean values of the ONSD for MRI (5.86 ± 0.66 mm) and transbulbar sonography (5.86 ± 0.71 mm) were identical. There was a high correlation between readers for measurements of the ONSD both for transbulbar sonography (CCC = 0.93) as well as for MRI (CCC = 0.9). Comparing the measurement values between transbulbar sonography and MRI, the correlation of ONSD values was good to moderate with a high dependency on image quality (CCC [0.31, 0.68]).

CONCLUSION

We were able to demonstrate that the ONSD values of transbulbar sonography in children correlate well with MRI if the relevant anatomic structures are depicted and the measuring points are set correctly. Based on the findings of our study, it seems necessary to define normal and cut-off values for the ONSD in children again to finally assess the clinical relevance of the ONSD as a noninvasive parameter for the evaluation of intracranial pressure.

Authors+Show Affiliations

Institut für Diagnostische und Interventionelle Radiologie und Kinderradiologie, Städtisches Klinikum München GmbH, Klinikum Schwabing.Institut für Diagnostische und Interventionelle Radiologie und Kinderradiologie, Städtisches Klinikum München GmbH, Klinikum Schwabing.Institut für Diagnostische und Interventionelle Radiologie und Kinderradiologie, Städtisches Klinikum München GmbH, Klinikum Schwabing.Institut für Medizinische Statistik und Epidemiologie, Technische Universität München.Kinderklinik und Poliklinik des Klinikums Rechts der Isar der Technischen Universität München, Kinderklinik Schwabing, Städtisches Klinikum München GmbH, Klinikum Schwabing.Institut für Diagnostische und Interventionelle Radiologie und Kinderradiologie, Städtisches Klinikum München GmbH, Klinikum Schwabing.Institut für Diagnostische und Interventionelle Radiologie und Kinderradiologie, Städtisches Klinikum München GmbH, Klinikum Schwabing.

Pub Type(s)

Journal Article

Language

eng

PubMed ID

21870318

Citation

Steinborn, M, et al. "Measurement of the Optic Nerve Sheath Diameter in Children: Comparison Between Transbulbar Sonography and Magnetic Resonance Imaging." Ultraschall in Der Medizin (Stuttgart, Germany : 1980), vol. 33, no. 6, 2012, pp. 569-573.
Steinborn M, Fiegler J, Ruedisser K, et al. Measurement of the Optic Nerve Sheath Diameter in Children: Comparison Between Transbulbar Sonography and Magnetic Resonance Imaging. Ultraschall Med. 2012;33(6):569-573.
Steinborn, M., Fiegler, J., Ruedisser, K., Hapfelmeier, A., Denne, C., Macdonald, E., & Hahn, H. (2012). Measurement of the Optic Nerve Sheath Diameter in Children: Comparison Between Transbulbar Sonography and Magnetic Resonance Imaging. Ultraschall in Der Medizin (Stuttgart, Germany : 1980), 33(6), 569-573. https://doi.org/10.1055/s-0031-1273491
Steinborn M, et al. Measurement of the Optic Nerve Sheath Diameter in Children: Comparison Between Transbulbar Sonography and Magnetic Resonance Imaging. Ultraschall Med. 2012;33(6):569-573. PubMed PMID: 21870318.
* Article titles in AMA citation format should be in sentence-case
TY - JOUR T1 - Measurement of the Optic Nerve Sheath Diameter in Children: Comparison Between Transbulbar Sonography and Magnetic Resonance Imaging. AU - Steinborn,M, AU - Fiegler,J, AU - Ruedisser,K, AU - Hapfelmeier,A, AU - Denne,C, AU - Macdonald,E, AU - Hahn,H, Y1 - 2011/08/25/ PY - 2011/8/27/pubmed PY - 2011/8/27/medline PY - 2011/8/27/entrez SP - 569 EP - 573 JF - Ultraschall in der Medizin (Stuttgart, Germany : 1980) JO - Ultraschall Med VL - 33 IS - 6 N2 - PURPOSE: To evaluate the accuracy of measurements of the optic nerve sheath diameter (ONSD) in children by comparing transbulbar sonography with magnetic resonance imaging (MRI). MATERIALS AND METHODS: 65 children (age: 3 mo- 17y; mean age: 11.3y) underwent MR imaging of the brain including a heavily T 2-w sequence of the orbit and transbulbar sonography using a 17 MHz linear array transducer. Measurements of the ONSD were performed retrospectively by two experienced readers and all images were evaluated regarding the image quality. Bland-Altman Plots were produced to assess the accuracy of measurements. The correlation between readers and between MR imaging and transbulbar sonography was calculated by the concordance correlation coefficient (CCC). RESULTS: Overall the mean values of the ONSD for MRI (5.86 ± 0.66 mm) and transbulbar sonography (5.86 ± 0.71 mm) were identical. There was a high correlation between readers for measurements of the ONSD both for transbulbar sonography (CCC = 0.93) as well as for MRI (CCC = 0.9). Comparing the measurement values between transbulbar sonography and MRI, the correlation of ONSD values was good to moderate with a high dependency on image quality (CCC [0.31, 0.68]). CONCLUSION: We were able to demonstrate that the ONSD values of transbulbar sonography in children correlate well with MRI if the relevant anatomic structures are depicted and the measuring points are set correctly. Based on the findings of our study, it seems necessary to define normal and cut-off values for the ONSD in children again to finally assess the clinical relevance of the ONSD as a noninvasive parameter for the evaluation of intracranial pressure. SN - 1438-8782 UR - https://www.unboundmedicine.com/medline/citation/21870318/Measurement_of_the_Optic_Nerve_Sheath_Diameter_in_Children:_Comparison_Between_Transbulbar_Sonography_and_Magnetic_Resonance_Imaging_ L2 - http://www.thieme-connect.com/DOI/DOI?10.1055/s-0031-1273491 DB - PRIME DP - Unbound Medicine ER -