Noninvasive CPAP with face mask: comparison among new air-entrainment masks and the Boussignac valve.
Respir Care. 2013 Feb; 58(2):305-12.RC

Abstract

BACKGROUND

The performances of 2 noninvasive CPAP systems (high flow and low flow air-entrainment masks) were compared to the Boussignac valve in 3 different scenarios.

METHODS

Scenario 1: pneumatic lung simulator with a tachypnea pattern (tidal volume 800 mL at 40 breaths/min). Scenario 2: Ten healthy subjects studied during tidal breaths and tachypnea. Scenario 3: Twenty ICU subjects enrolled for a noninvasive CPAP session. Differences between set and effective CPAP level and F(IO(2)), as well as the lowest airway pressure and the pressure swing around the imposed CPAP level, were analyzed. The lowest airway pressure and swing were correlated to the pressure-time product (area of the airway pressure curve below the CPAP level) measured with the simulator. P(aO(2)) was a subject's further performance index.

RESULTS

Lung simulator: Boussignac F(IO(2)) was 0.54, even if supplied with pure oxygen. The air-entrainment masks had higher swing than the Boussignac (P = .007). Pressure-time product correlated better with pressure swing (Spearman correlation coefficient [ρ] = 0.97) than with lowest airway pressure (ρ = 0.92). In healthy subjects, the high-flow air-entrainment mask showed lower difference between set and effective F(IO(2)) (P < .001), and lowest airway pressure (P < .001), compared to the Boussignac valve. In all measurements the Boussignac valve showed higher than imposed CPAP level (P < .001). In ICU subjects the high-flow mask had lower swing than the Boussignac valve (P = .03) with similar P(aO(2)) increase.

CONCLUSIONS

High-flow air-entrainment mask showed the best performance in human subjects. During high flow demand, the Boussignac valve delivered lower than expected F(IO(2)) and showed higher dynamic hyper-pressurization than the air-entrainment masks.

Authors+Show Affiliations

Dipartimento di Fisiopatologia Medico-Chirurgica e dei Trapianti, Università degli Studi di Milano, Polo Universitario San Paolo, Milan, Italy. giovanni.mistraletti@unimi.itNo affiliation info availableNo affiliation info availableNo affiliation info availableNo affiliation info availableNo affiliation info availableNo affiliation info availableNo affiliation info availableNo affiliation info availableNo affiliation info available

Pub Type(s)

Comparative Study
Journal Article

Language

eng

PubMed ID

22709435

Citation

Mistraletti, Giovanni, et al. "Noninvasive CPAP With Face Mask: Comparison Among New Air-entrainment Masks and the Boussignac Valve." Respiratory Care, vol. 58, no. 2, 2013, pp. 305-12.
Mistraletti G, Giacomini M, Sabbatini G, et al. Noninvasive CPAP with face mask: comparison among new air-entrainment masks and the Boussignac valve. Respir Care. 2013;58(2):305-12.
Mistraletti, G., Giacomini, M., Sabbatini, G., Pinciroli, R., Mantovani, E. S., Umbrello, M., Palmisano, D., Formenti, P., Destrebecq, A. L., & Iapichino, G. (2013). Noninvasive CPAP with face mask: comparison among new air-entrainment masks and the Boussignac valve. Respiratory Care, 58(2), 305-12. https://doi.org/10.4187/respcare.01598
Mistraletti G, et al. Noninvasive CPAP With Face Mask: Comparison Among New Air-entrainment Masks and the Boussignac Valve. Respir Care. 2013;58(2):305-12. PubMed PMID: 22709435.
* Article titles in AMA citation format should be in sentence-case
TY - JOUR T1 - Noninvasive CPAP with face mask: comparison among new air-entrainment masks and the Boussignac valve. AU - Mistraletti,Giovanni, AU - Giacomini,Matteo, AU - Sabbatini,Giovanni, AU - Pinciroli,Riccardo, AU - Mantovani,Elena S, AU - Umbrello,Michele, AU - Palmisano,Debora, AU - Formenti,Paolo, AU - Destrebecq,Anne L L, AU - Iapichino,Gaetano, PY - 2012/6/20/entrez PY - 2012/6/20/pubmed PY - 2014/1/28/medline SP - 305 EP - 12 JF - Respiratory care JO - Respir Care VL - 58 IS - 2 N2 - BACKGROUND: The performances of 2 noninvasive CPAP systems (high flow and low flow air-entrainment masks) were compared to the Boussignac valve in 3 different scenarios. METHODS: Scenario 1: pneumatic lung simulator with a tachypnea pattern (tidal volume 800 mL at 40 breaths/min). Scenario 2: Ten healthy subjects studied during tidal breaths and tachypnea. Scenario 3: Twenty ICU subjects enrolled for a noninvasive CPAP session. Differences between set and effective CPAP level and F(IO(2)), as well as the lowest airway pressure and the pressure swing around the imposed CPAP level, were analyzed. The lowest airway pressure and swing were correlated to the pressure-time product (area of the airway pressure curve below the CPAP level) measured with the simulator. P(aO(2)) was a subject's further performance index. RESULTS: Lung simulator: Boussignac F(IO(2)) was 0.54, even if supplied with pure oxygen. The air-entrainment masks had higher swing than the Boussignac (P = .007). Pressure-time product correlated better with pressure swing (Spearman correlation coefficient [ρ] = 0.97) than with lowest airway pressure (ρ = 0.92). In healthy subjects, the high-flow air-entrainment mask showed lower difference between set and effective F(IO(2)) (P < .001), and lowest airway pressure (P < .001), compared to the Boussignac valve. In all measurements the Boussignac valve showed higher than imposed CPAP level (P < .001). In ICU subjects the high-flow mask had lower swing than the Boussignac valve (P = .03) with similar P(aO(2)) increase. CONCLUSIONS: High-flow air-entrainment mask showed the best performance in human subjects. During high flow demand, the Boussignac valve delivered lower than expected F(IO(2)) and showed higher dynamic hyper-pressurization than the air-entrainment masks. SN - 1943-3654 UR - https://www.unboundmedicine.com/medline/citation/22709435/Noninvasive_CPAP_with_face_mask:_comparison_among_new_air_entrainment_masks_and_the_Boussignac_valve_ DB - PRIME DP - Unbound Medicine ER -