QuEChERS-Based Extraction Procedures for the Analysis of Bisphenols S and A in Breast Milk Samples by LC-QqQ-MS.
J AOAC Int. 2018 Oct 17 [Online ahead of print]JA

Abstract

Different extraction and clean-up protocols, based on either the dispersive solid-phase extraction (d-SPE) and the quick, easy, cheap, effective, rugged, and safe (QuEChERS) approach, were optimized and compared for the determination of selected bisphenols such as bisphenol A (BPA) and bisphenol S (BPS) in breast milk samples. QuEChERS-based sample preparation procedure was optimized by evaluation of different clean-up sorbents including zirconium-based sorbents (Z-Sep and Z-Sep Plus) and primary secondary amine. Negligible matrix interference was observed for most of the analytes due to application of 45 mg Z-Sep or 45 mg Z-Sep + 5 mg Z-Sep Plus sorbents in d-SPE clean-up step. Acceptable analytical performance for the BPA and BPS was observed with recoveries percentage in the range of 70-92% and RSD less than 15%. The pilot study was performed by applying HPLC with diode-array detection, and optimized procedures were transferred to the LC triple-quadrupole MS system. LC with electrospray ionization and tandem MS operating in the multiple reaction monitoring mode provide high sensitivity and selectivity for trace analysis. The LODs were 0.10 and 0.54 ng/mL, and the LOQs were 0.20 and 1.35 ng/mL for BPS and BPA, respectively. The developed procedures were evaluated in terms for material collection obtained from women (inhabitants of Lublin, Poland). In 20 samples, bisphenol residues were detected at concentrations from 0.09 to 11.56 ng/mL.

Authors+Show Affiliations

Medical University in Lublin, Department of Physical Chemistry, Chair of Chemistry, 4A Chodźki St, Lublin 20-093, Poland.Medical University of Lublin, Department of Forensic Medicine, 8b Jaczewskiego St, Lublin 20-090, Poland.Medical University of Lublin, Department of Forensic Medicine, 8b Jaczewskiego St, Lublin 20-090, Poland.Medical University of Lublin, Department of Forensic Medicine, 8b Jaczewskiego St, Lublin 20-090, Poland.

Pub Type(s)

Journal Article

Language

eng

PubMed ID

30333076

Citation

Tuzimski, Tomasz, et al. "QuEChERS-Based Extraction Procedures for the Analysis of Bisphenols S and a in Breast Milk Samples By LC-QqQ-MS." Journal of AOAC International, 2018.
Tuzimski T, Pieniążek D, Buszewicz G, et al. QuEChERS-Based Extraction Procedures for the Analysis of Bisphenols S and A in Breast Milk Samples by LC-QqQ-MS. J AOAC Int. 2018.
Tuzimski, T., Pieniążek, D., Buszewicz, G., & Teresiński, G. (2018). QuEChERS-Based Extraction Procedures for the Analysis of Bisphenols S and A in Breast Milk Samples by LC-QqQ-MS. Journal of AOAC International. https://doi.org/10.5740/jaoacint.18-0297
Tuzimski T, et al. QuEChERS-Based Extraction Procedures for the Analysis of Bisphenols S and a in Breast Milk Samples By LC-QqQ-MS. J AOAC Int. 2018 Oct 17; PubMed PMID: 30333076.
* Article titles in AMA citation format should be in sentence-case
TY - JOUR T1 - QuEChERS-Based Extraction Procedures for the Analysis of Bisphenols S and A in Breast Milk Samples by LC-QqQ-MS. AU - Tuzimski,Tomasz, AU - Pieniążek,Dominika, AU - Buszewicz,Grzegorz, AU - Teresiński,Grzegorz, Y1 - 2018/10/17/ PY - 2018/10/20/pubmed PY - 2018/10/20/medline PY - 2018/10/19/entrez JF - Journal of AOAC International JO - J AOAC Int N2 - Different extraction and clean-up protocols, based on either the dispersive solid-phase extraction (d-SPE) and the quick, easy, cheap, effective, rugged, and safe (QuEChERS) approach, were optimized and compared for the determination of selected bisphenols such as bisphenol A (BPA) and bisphenol S (BPS) in breast milk samples. QuEChERS-based sample preparation procedure was optimized by evaluation of different clean-up sorbents including zirconium-based sorbents (Z-Sep and Z-Sep Plus) and primary secondary amine. Negligible matrix interference was observed for most of the analytes due to application of 45 mg Z-Sep or 45 mg Z-Sep + 5 mg Z-Sep Plus sorbents in d-SPE clean-up step. Acceptable analytical performance for the BPA and BPS was observed with recoveries percentage in the range of 70-92% and RSD less than 15%. The pilot study was performed by applying HPLC with diode-array detection, and optimized procedures were transferred to the LC triple-quadrupole MS system. LC with electrospray ionization and tandem MS operating in the multiple reaction monitoring mode provide high sensitivity and selectivity for trace analysis. The LODs were 0.10 and 0.54 ng/mL, and the LOQs were 0.20 and 1.35 ng/mL for BPS and BPA, respectively. The developed procedures were evaluated in terms for material collection obtained from women (inhabitants of Lublin, Poland). In 20 samples, bisphenol residues were detected at concentrations from 0.09 to 11.56 ng/mL. SN - 1944-7922 UR - https://www.unboundmedicine.com/medline/citation/30333076/QuEChERS_Based_Extraction_Procedures_for_the_Analysis_of_Bisphenols_S_and_A_in_Breast_Milk_Samples_by_LC_QqQ_MS_ DB - PRIME DP - Unbound Medicine ER -