Tags

Type your tag names separated by a space and hit enter

Virtiol - Simulation of Quality of Vision with Multifocal and Edof Intraocular Lenses.
Cesk Slov Oftalmol 2019; 74(6):219-225CS

Abstract

Pupose: To experimentally compare the visual acuity and the subjective perception of different types of multifocal intraocular lenses (IOL) using a VirtIOL device/simulator in a group of volunteers with artephakia.

MATERIAL AND METHODS

This was an experimental study involving a total of 20 volunteers with artephakia (35 eyes). Each volunteer rated 5 types of IOLs, 4 presbyopia-correcting IOLs - WIOL-CF, Tecnis Symphony ZXR00, Acrysof IQ PanOptix TFNT00, M-flex 630 F, and as a reference lens, we used the monofocal IOL Acrysof SA60AT. The corrected distance visual acuity (CDVA), distance corrected intermediated visual acuity (DCIVA) and distance corrected near visual acuity (DCNVA) were measured. Additionally, volunteers evaluated the quality of vision under normal or changed lighting conditions, and ranked IOL on scale from 1 to 5.

RESULTS

The CDVA evaluated using the VirtIOL device was very good for all tested IOLs (0.04-0.09 log MAR) with minimum differences. However, CDVA without simulator (-0.01 logMAR) was statistically significantly better in all cases. DCIVA was also very similar in each of the investigated IOLs, surprisingly even with monofocal IOL (0.21-0.23 logMAR), without using simulator the DCIVA was statistically significantly worse (0.36 logMAR). The DCNVA was the best for PanOptix intraocular lens (0.22 logMAR); M-flex, Symphony and WIOL-CF lenses had comparable results (0.31-0.34 logMAR). Again, surprisingly similar results were obtained with the use of monofocal IOL (0.36 ± 0.14). Subjective perception of vision through the IOLs was best rated for the monofocal control IOL, whereas Symphony, WIOL-CF and M-flex did not show any statistically significant difference either with or without glare. All tested IOLs were statistically significantly better if compared to PanOptix with or without glare.

CONCLUSION

Simulation of vision through IOLs using VirtIOL simulator allows to compare different models of multifocal IOLs from the viewpoint of visual acuity and subjective perception. However, some caution should be exercised when evaluating the results, given that in our experiments, the monofocal IOL achieved relatively good results at near distance, which does not correspond to clinical experience. On the contrary, from the comparison of the results of CDVA without and with VirtIOL, it is obvious that visual acuity is slightly adversely affected by added optics.

Authors

No affiliation info availableNo affiliation info availableNo affiliation info availableNo affiliation info availableNo affiliation info available

Pub Type(s)

Journal Article

Language

eng

PubMed ID

31238689

Citation

Studený, P, et al. "Virtiol - Simulation of Quality of Vision With Multifocal and Edof Intraocular Lenses." Ceska a Slovenska Oftalmologie : Casopis Ceske Oftalmologicke Spolecnosti a Slovenske Oftalmologicke Spolecnosti, vol. 74, no. 6, 2019, pp. 219-225.
Studený P, Hlaváček J, Chaloupka R, et al. Virtiol - Simulation of Quality of Vision with Multifocal and Edof Intraocular Lenses. Cesk Slov Oftalmol. 2019;74(6):219-225.
Studený, P., Hlaváček, J., Chaloupka, R., Veselý, Ľ., & Baxant, A. D. (2019). Virtiol - Simulation of Quality of Vision with Multifocal and Edof Intraocular Lenses. Ceska a Slovenska Oftalmologie : Casopis Ceske Oftalmologicke Spolecnosti a Slovenske Oftalmologicke Spolecnosti, 74(6), pp. 219-225. doi:10.31348/2018/6/1.
Studený P, et al. Virtiol - Simulation of Quality of Vision With Multifocal and Edof Intraocular Lenses. Cesk Slov Oftalmol. 2019;74(6):219-225. PubMed PMID: 31238689.
* Article titles in AMA citation format should be in sentence-case
TY - JOUR T1 - Virtiol - Simulation of Quality of Vision with Multifocal and Edof Intraocular Lenses. AU - Studený,P, AU - Hlaváček,J, AU - Chaloupka,R, AU - Veselý,Ľ, AU - Baxant,A D, PY - 2019/6/27/entrez PY - 2019/6/27/pubmed PY - 2019/8/27/medline KW - vision simulator KW - VirtIOL KW - multifocal intraocular lens KW - presbyopia correction SP - 219 EP - 225 JF - Ceska a slovenska oftalmologie : casopis Ceske oftalmologicke spolecnosti a Slovenske oftalmologicke spolecnosti JO - Cesk Slov Oftalmol VL - 74 IS - 6 N2 - : Pupose: To experimentally compare the visual acuity and the subjective perception of different types of multifocal intraocular lenses (IOL) using a VirtIOL device/simulator in a group of volunteers with artephakia. MATERIAL AND METHODS: This was an experimental study involving a total of 20 volunteers with artephakia (35 eyes). Each volunteer rated 5 types of IOLs, 4 presbyopia-correcting IOLs - WIOL-CF, Tecnis Symphony ZXR00, Acrysof IQ PanOptix TFNT00, M-flex 630 F, and as a reference lens, we used the monofocal IOL Acrysof SA60AT. The corrected distance visual acuity (CDVA), distance corrected intermediated visual acuity (DCIVA) and distance corrected near visual acuity (DCNVA) were measured. Additionally, volunteers evaluated the quality of vision under normal or changed lighting conditions, and ranked IOL on scale from 1 to 5. RESULTS: The CDVA evaluated using the VirtIOL device was very good for all tested IOLs (0.04-0.09 log MAR) with minimum differences. However, CDVA without simulator (-0.01 logMAR) was statistically significantly better in all cases. DCIVA was also very similar in each of the investigated IOLs, surprisingly even with monofocal IOL (0.21-0.23 logMAR), without using simulator the DCIVA was statistically significantly worse (0.36 logMAR). The DCNVA was the best for PanOptix intraocular lens (0.22 logMAR); M-flex, Symphony and WIOL-CF lenses had comparable results (0.31-0.34 logMAR). Again, surprisingly similar results were obtained with the use of monofocal IOL (0.36 ± 0.14). Subjective perception of vision through the IOLs was best rated for the monofocal control IOL, whereas Symphony, WIOL-CF and M-flex did not show any statistically significant difference either with or without glare. All tested IOLs were statistically significantly better if compared to PanOptix with or without glare. CONCLUSION: Simulation of vision through IOLs using VirtIOL simulator allows to compare different models of multifocal IOLs from the viewpoint of visual acuity and subjective perception. However, some caution should be exercised when evaluating the results, given that in our experiments, the monofocal IOL achieved relatively good results at near distance, which does not correspond to clinical experience. On the contrary, from the comparison of the results of CDVA without and with VirtIOL, it is obvious that visual acuity is slightly adversely affected by added optics. SN - 1211-9059 UR - https://www.unboundmedicine.com/medline/citation/31238689/Virtiol_-_Simulation_of_Quality_of_Vision_with_Multifocal_and_Edof_Intraocular_Lenses L2 - https://www.prolekare.cz/linkout/112502 DB - PRIME DP - Unbound Medicine ER -