Validation of the InBody BP170 oscillometric home blood pressure monitor in general population according to the Association for the Advancement of Medical Instrumentation/European Society of Hypertension/International Organization for Standardization Universal Standard.

Abstract

OBJECTIVE

To evaluate the accuracy of the InBody BP170 oscillometric upper-arm home blood pressure (BP) monitor in general population according to the Association for the Advancement of Medical Instrumentation/European Society of Hypertension/International Organization for Standardization (AAMI/ESH/ISO) Universal Standard (ISO 81060-2:2018).

METHODS

Subjects were recruited to fulfill the age, gender, BP and cuff distribution criteria of the AAMI/ESH/ISO Universal Standard in general population using the same arm sequential BP measurement method. Two cuffs of the test device were used for arm circumference 22-32 (medium) and 32-42 cm (large).

RESULTS

One-hundred five subjects were recruited and 88 were analyzed. For validation Criterion 1, the mean ± SD of the differences between the test device and reference BP readings was 0.9 ± 6.9/0.5 ± 5.5 mmHg (systolic/diastolic). For Criterion 2, the SD of the averaged BP differences between the test device and reference BP per subject was 5.77/4.80 mmHg (systolic/diastolic).

CONCLUSION

The InBody BP170 oscillometric home BP monitor fulfilled all the requirements of the AAMI/ESH/ISO Universal Standard (ISO 81060-2:2018) in general population and can be recommended for clinical use.

Authors+Show Affiliations

Hypertension Center STRIDE-7, National and Kapodistrian University of Athens, School of Medicine, Third Department of Medicine, Sotiria Hospital, Athens, Greece.No affiliation info availableNo affiliation info availableNo affiliation info availableNo affiliation info availableNo affiliation info availableNo affiliation info available

Pub Type(s)

Journal Article

Language

eng

PubMed ID

31633518

Citation

Kollias, Anastasios, et al. "Validation of the InBody BP170 Oscillometric Home Blood Pressure Monitor in General Population According to the Association for the Advancement of Medical Instrumentation/European Society of Hypertension/International Organization for Standardization Universal Standard." Blood Pressure Monitoring, 2019.
Kollias A, Anagnostopoulos I, Gravvani A, et al. Validation of the InBody BP170 oscillometric home blood pressure monitor in general population according to the Association for the Advancement of Medical Instrumentation/European Society of Hypertension/International Organization for Standardization Universal Standard. Blood Press Monit. 2019.
Kollias, A., Anagnostopoulos, I., Gravvani, A., Stambolliu, E., Bountzona, I., Menti, A., & Stergiou, G. S. (2019). Validation of the InBody BP170 oscillometric home blood pressure monitor in general population according to the Association for the Advancement of Medical Instrumentation/European Society of Hypertension/International Organization for Standardization Universal Standard. Blood Pressure Monitoring, doi:10.1097/MBP.0000000000000410.
Kollias A, et al. Validation of the InBody BP170 Oscillometric Home Blood Pressure Monitor in General Population According to the Association for the Advancement of Medical Instrumentation/European Society of Hypertension/International Organization for Standardization Universal Standard. Blood Press Monit. 2019 Oct 15; PubMed PMID: 31633518.
* Article titles in AMA citation format should be in sentence-case
TY - JOUR T1 - Validation of the InBody BP170 oscillometric home blood pressure monitor in general population according to the Association for the Advancement of Medical Instrumentation/European Society of Hypertension/International Organization for Standardization Universal Standard. AU - Kollias,Anastasios, AU - Anagnostopoulos,Ioannis, AU - Gravvani,Areti, AU - Stambolliu,Emelina, AU - Bountzona,Ioanna, AU - Menti,Ariadni, AU - Stergiou,George S, Y1 - 2019/10/15/ PY - 2019/10/22/entrez JF - Blood pressure monitoring JO - Blood Press Monit N2 - OBJECTIVE: To evaluate the accuracy of the InBody BP170 oscillometric upper-arm home blood pressure (BP) monitor in general population according to the Association for the Advancement of Medical Instrumentation/European Society of Hypertension/International Organization for Standardization (AAMI/ESH/ISO) Universal Standard (ISO 81060-2:2018). METHODS: Subjects were recruited to fulfill the age, gender, BP and cuff distribution criteria of the AAMI/ESH/ISO Universal Standard in general population using the same arm sequential BP measurement method. Two cuffs of the test device were used for arm circumference 22-32 (medium) and 32-42 cm (large). RESULTS: One-hundred five subjects were recruited and 88 were analyzed. For validation Criterion 1, the mean ± SD of the differences between the test device and reference BP readings was 0.9 ± 6.9/0.5 ± 5.5 mmHg (systolic/diastolic). For Criterion 2, the SD of the averaged BP differences between the test device and reference BP per subject was 5.77/4.80 mmHg (systolic/diastolic). CONCLUSION: The InBody BP170 oscillometric home BP monitor fulfilled all the requirements of the AAMI/ESH/ISO Universal Standard (ISO 81060-2:2018) in general population and can be recommended for clinical use. SN - 1473-5725 UR - https://www.unboundmedicine.com/medline/citation/31633518/Validation_of_the_InBody_BP170_oscillometric_home_blood_pressure_monitor_in_general_population_according_to_the_Association_for_the_Advancement_of_Medical_Instrumentation/European_Society_of_Hypertension/International_Organization_for_Standardization_Universal_Standard L2 - http://dx.doi.org/10.1097/MBP.0000000000000410 DB - PRIME DP - Unbound Medicine ER -